  • Prospective adaptive navigator correction for breath-hold MR coronary angiography.

    McConnell MV, Khasgiwala VC, Savord BJ, Chen MH, Chuang ML, Edelman RR, Manning WJ. Magn Reson Med. 1997; 37 (1): 148-52

    Current MR coronary angiography (MRCA) methods use breath-holding to minimize respiratory motion. A major limitation to this technique is misregistration between imaging slices due to breath-hold variability. Prospective adaptive correction of image location using real-time navigator measurement of diaphragm position is a potential method for improving slice registration in breath-hold MRCA. Ten subjects underwent MRCA using an ECG-gated, fat-suppressed, segmented k-space, gradient-echo sequence. Transverse and coronal images were acquired using standard breath-holding with and without prospective navigator correction. Breath-hold MRCA with prospective navigator correction resulted in a 47% reduction in craniocaudal slice registration error compared to standard breath-holding (0.9 +/- 0.2 mm versus 1.7 +/- 0.4 mm, P = 0.04). Prospective adaptive navigator correction of image location significantly improves slice registration for breath-hold MRCA and is a promising motion correction technique for cardiac MR.