Paul J. Wang, MD

Publication Details

  • Inappropriate shock: A failure of SVT discriminators in a dual chamber ICD? PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY Al-Ahmad, A., Tsiperfal, A., Hsia, H. H., Wang, P. J. 2006; 29 (12): 1413-1415

    Abstract:

    Inappropriate shock remains a major issue in patients with implantable cardioverter defibrillators. We examine an inappropriate shock delivered in a patient with atrial tachycardia that conducted 1:1. We reconstruct the device algorithms that led to therapy delivery and discuss programming changes that could be considered.

    View details for Web of Science ID 000243530800015

    View details for PubMedID 17201851