R Brooke Jeffrey

Publication Details

  • QUANTITATIVE-EVALUATION OF PANCREATIC ENHANCEMENT DURING DUAL-PHASE HELICAL CT RADIOLOGY HOLLETT, M. D., Jorgensen, M. J., Jeffrey, R. B. 1995; 195 (2): 359-361

    Abstract:

    To determine the improvement in pancreatic enhancement at helical computed tomography (CT) performed with an early delay after administration of contrast material compared with that performed with a standard delay.Dual-phase helical CT of the abdomen was performed in 120 patients with a 150-mL bolus of contrast material infused at 5 mL/sec. Early and standard delayed scanning was performed beginning at 20 seconds and 49-71 seconds, respectively. Regions of interest were measured in the head, body, and tail of the pancreas in 92 patients. The difference in enhancement between early and standard delayed scanning was calculated.Mean pancreatic enhancement was 82 HU +/- 3 (standard error) with an early delay, whereas enhancement on standard delay scans was 62 HU +/- 2 (P < .001). An improvement in enhancement greater than 10 HU was attained in 66 of 92 cases (72%).Pancreatic enhancement at helical CT with an early delay after contrast material administration is often significantly greater than the enhancement seen with a standard delay when a monophasic, rapidly infused bolus of contrast material is used.

    View details for Web of Science ID A1995QU71700013

    View details for PubMedID 7724753

Stanford Medicine Resources:

Footer Links: